www.ascensia.ca
Meter Specifi cations:
Test Sample: Whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 µL
Measuring Range: 0.6–33.3 mmol/L
Measuring Time: 5 seconds with Ascensia CONTOUR Test
Strips; 15 seconds with Ascensia
M
ICROFILL Test Strips
Memory Feature: Stores most recent 480 test results
Battery Type: Two 3-volt lithium batteries
(DL2032 or CR2032)
Battery Life: Approximately 1000 tests (1 yr. average use)
Operating Temperature Range: 5°C–45°C
Humidity: 1093% RH
Dimensions: 77 mm (H) x 57 mm (W) x 19 mm (T)
Weight: 47.5 grams (or 1.7 ounces)
Alarm: “Beeps” whenever a test strip is inserted into or
removed from the meter, when the test strip is lled with
blood or control solution, and when a test result appears
in the display. Two beeps will sound for an error. Twenty
beeps will sound when a programmed alarm goes off.